Basic Info.
Type of Body Fluid-Processing Device
Ingredients Blood Transfusion Device
Back up Battery
30mintues
Transport Package
Wooden Case
Specification
380mm*400*1380mm (L*W*H)
Product Description
Product description:
Technical Parameters of Hemodialysis Machine ZY-T2008-B
Main Performance
Self-checking function
Carbonate dialysis
Double-needle dialysis
Liquid level detector
Bubble detector
Blood leakage detector
Temperature and conductivity monitoring
Arterial pressure, venous pressure and transmembrane pressure monitoring
Rolling blood pump
Heparin pump
Dehydration amount controlled by capacity
Automatic disinfection and cleaning program
Stand-by power of blood pump in case of power failure
Information display function of display screen
Technical Data
General Data
Dimensions: 1380 x 380 x 400 mm (H x W x D)
Net weight approximately: 88 KG
Package dimensions: 570X660X1581mm (L x W x H)
Gross weight approximately: 95KG
Water Supply
Water inlet pressure: 1.5 - 6.0 bar
Water inlet temperature: 5°C - 32°C
Max. drain height: 1m
Concentrate Supply
Supply pressure 0 to 2m suction height
Electrical Data
Power supply: AC220V±10%, 47 - 63HZ
Current consumption max.: 10A
Extracorporeal Circuit
Arterial Pressure Monitoring
Display range: -380 mmHg to +400 mmHg
Accuracy: ±10 mmHg
Resolution: 1 mmHg
Venous Pressure Monitoring
Display range: -180 mmHg to +600 mmHg
Accuracy: ±10 mmHg
Resolution: 1 mmHg
Transmembrane Pressure Monitoring
Display range: -180 mmHg to +600 mmHg
Accuracy: ±20 mmHg
Resolution: 1 mmHg
Arterial Blood Pump
Blood flow range: 30 to 600ml/min in 8 mm bloodline systems
Accuracy: ±10%
Air Bubble Pump : By ultrasound transmission, with additional optical monitoring in venous clamp.
Heparin Pump
Delivery range: 0ml/h to 10ml/h
Bolus function max.: 20 mL per bolus
Syringe size: 10, 20, 30, 50ml
Dialysis Fluid Circuit
Dialysis Fluid Flow Range : Selectable 300 - 800 mL/min
Dialysis Fluid Temperature
Selectable: 34.0°C to 40.0°C
Dialysis Fluid Conductivity
Range: 12 to 16 mS/cm (25°C)
Accuracy: ±0.1 mS/cm
Acid Concentration Dialysis Fluid
Default mixing ratio: 1 + 34 (others possible)
Range: 125 to 150 mmol/L
Bicarbonate Concentration Dialysis Fluid
Default mixing ratio: 1 + 27.6 (others possible)
Range: -8 to +8 mmol/L bicarbonate
Ultrafiltration
UF rate: 0ml/h - 4000ml/h
Accuracy: ±1%
Allowed dialyzer UF factor unlimited
Parameter displayed: UF goal, UF time, UF rate, UF volume
Blood Leak Detector
Sensitivity: ≤0.35 mL blood/min (HCT = 32)
KT/V Calculation
Disinfection and Cleaning Programs
Rinse: Temperature/flow 30 to 60ºC/ 300 to 800ml/min
Hot Rinse (recirculation): Temperature/flow 30 to 85ºC/ 300 to 800ml/min
Hot disinfection with chemical: Temperature/flow 30 to 85ºC/ 300 to 800ml/min
Various program combinations selectable
Optional Technical Data
Blood Pressure Monitor (BPM) (Optional Function)
Display range: Systole: 40 - 280 mmHg, Diastole: 40 - 280 mmHg
Accuracy: 1 mmHg
Central Delivery System for Acid Conc.
Supply pressure 0 to 100 mbar; 1m suction height with Central delivery system: 0 - 500 mbar
Endotoxin Filter - Dialysis Fluid Filter System (Optional Function)
Balancing accuracy: ±0.1% of dialysate flow
Bicarbonate Dry Concentrate
Bi-cart (Optional Function)
View More