Product Parameters
Specification of CWM-201A Anesthesia
1.Feature
Integrated breathing circuit with APLvalve,excellent tightness.
Replacing drive bellow is NOT required when operating
Adult/pediatric anesthesia.
Applied with 1 high-precision vaporizers.(Selectatec).
10.1 inch screen.
Real-time detection forloops,Paw, BPM, VT and compliance. Comprehensive monitoring capabilties include ventilation parameters monitoring,preset parameters, alarm indication and breathing waveform monitoring.
ACGO available to connect with T/Bain circuit for
open operation.(Optional)
2.Physical specifcations :
Dimensions:70*85*140cm Weight:65kg
Casters:Front wheels with lock
LCD Screen:10.1 inch
Waveforms of P-T,F-T,VT
3.Working specifications:
Power:AC 110V~240V,50Hz/60Hz
Battery Power : ≥180 mins
Pipeline supply
Gas configuration:Oz,N2O,Air(Optional)
Pipeline input range:280-600kPa(40~85psi)
4.Ventilator Specifications :
Patient range :Adult ,Peadiatric
Modes of ventilation: VCV, SIMV, MANUAL, STAND-BY
5.Ventlator parameter
Working type:Electronic control,pneumatic driven
Tidal Volume range:20-1500ml, Pressure range:5-60 cmH2O
Rispriation Rate Range:1-100bpm(140bpm under SIMV),
l:E Range:4:1-1:8
Ptrigger Range:-20-20cmH2O,Minute volume range:>18L/Min, SIGH range: 0-5/100
6.System Standard
Hypoxic guard system:N2O cut-off valve,O2 concentration >25%
Safety valve :<12.5kPa
O2 Flush:25-75L/min
Flowmeter:4tubes,cascade 0-15L/min O2,0-15LminN2O,0-15Lmin Air(Optional)
Volume of CO2 absorber:2L
Vaporizer:1 Selectetac,maximum2mounts(Isoflurane/Enfurane/Sevoflurane/Halothane)
Monitoring:VT,MV,BPMPaw, Lung complince,FiOz Optional)
7.Alarm Limits:
Audio/Visual:No idal volume,MV,Paw,FIO2(Optional),Airfailure,AC powerfailure,Battry low
Pre-setting functions:
Language:CHINESE,ENGLISHetc.
Self-test:Automatic
8.Optional features
ACGO, Yoke system (Pin-index), AGSS
The anesthesia machine is a precious and precise instrument with a high utilization rate. It should be shared by special management, that is, it must be kept by a special person and stored in a dry ventilated and dark place. Do not put it in the same room with corrosive chemical reagents to ensure the spare quantity of various consumables. The pipeline after each use is cleaned and stored in time and kept in reserve to keep the appearance of the anesthesia machine clean.
The disinfection of anesthesia machine refers to the disinfection of the airway system of the anesthesia machine, and the host is generally only cleaned. The internal gas path of the main body of the anesthesia machine cannot be removed. Only the pipes of the patient's inhalation and exhalation circuit can be removed for cleaning and disinfection.
The main cleaning of the main body of the anesthesia machine is dehumidification and dust removal. If there is dust, it can be removed by a vacuum cleaner. The outside of the host can be cleaned with a damp cotton cloth, and the room can be naturally irradiated with ultraviolet rays or cleaned in a clean operating room.